The First Heart Sound Amplitude Using Digital Esophageal Stethoscope System is Proportional to The Change in Cardiac Output

Shin, Young Duck and Yim, Kyoung Hoon and Park, Sang Hi and Jeon, Yong Wook and Bae, Jin Ho Bae and Lee, Tae Soo and Kim, Myoung Hwan and Choi, Young Jin (1969) The First Heart Sound Amplitude Using Digital Esophageal Stethoscope System is Proportional to The Change in Cardiac Output. Pakistan Journal of Medical Sciences, 30 (2). ISSN 1681-715X

[thumbnail of 4383-20813-1-PB.pdf] Text
4383-20813-1-PB.pdf - Published Version

Download (1MB)

Abstract

Objective: The use of an esophageal stethoscope is a basic heart sounds monitoring procedure performed in patients under general anesthesia. As the size of the first heart sound can express the left ventricle function, its correlation with cardiac output should be investigated. The aim of this study was to investigate the effects of cardiac output (CO) on the first heart sound (S1) amplitude.

Methods: Six male beagles were chosen. The S1 was obtained with the newly developed esophageal stethoscope system. CO was measured using NICOM, a non-invasive CO measuring device. Ephedrine and beta blockers were administered to the subjects to compare changes in figures, and the change from using an inhalation anesthetic was also compared.

Results: The S1 amplitude displayed positive correlation with the change rate of CO (r = 0.935, p < 0.001). The heart rate measured using the esophageal stethoscope and ECG showed considerably close figures through the Bland-Altman plot and showed a high positive correlation (r = 0.988, p < 0.001).

Conclusion: In beagles, the amplitude of S1 had a significant correlation with changes in CO in a variety of situations.